IsoVu Isolated Probes
IsoVu® probes are the right tool for today’s demanding power measurement challenges given their industry leading 1 GHz bandwidth, 160 dB or 100 Million to 1 common mode rejection, 60 kV common mode voltage, large ± 2500 V differential range and superior probe loading.Learn how you can use IsoVu technology to make impossible measurements such as high side Vgs.
Model |
Bandwidth |
Differential Voltage |
Common Mode Voltage |
Common Mode Rejection Ratio |
Fiber Cable Length |
|
TIVH02 |
200 MHz |
± 2500 V
|
≤ 60 kV |
DC: > 160 dB
100 MHz: 100 dB
200 MHz: 100 dB
|
3 meters |
|
TIVH02L |
200 MHz |
± 2500 V
|
≤ 60 kV |
DC: > 160 dB
100 MHz: 100 dB
200 MHz: 100 dB
|
10 meters |
|
TIVH05 |
500 MHz |
± 2500 V |
≤ 60 kV |
DC: > 160 dB
100 MHz: 100 dB
500 MHz: 80 dB
|
3 meters |
|
TIVH05L |
500 MHz |
± 2500 V |
≤ 60 kV |
DC: > 160 dB
100 MHz: 100 dB
500 MHz: 80 dB
|
10 meters |
|
TIVM1 |
1 GHz |
± 50 V |
≤ 60 kV
|
DC: > 160 dB
100 MHz: 120 dB
1 GHz: 80 dB
|
3 meters |
|
TIVH08 |
800 MHz |
± 2500 V |
≤ 60 kV |
DC: > 160 dB
100 MHz: 100 dB
800 MHz: 75 dB
|
3 meters |
|
TIVM1L |
1 GHz |
± 50 V |
≤ 60 kV
|
DC: > 160 dB
100 MHz: 120 dB
1 GHz: 80 dB
|
10 meters |
|
TIVH08L |
800 MHz |
± 2500 V |
≤ 60 kV |
DC: > 160 dB
100 MHz: 100 dB
800 MHz: 75 dB
|
10 meters |
|
High Bandwidth Optical Probes
The DPO7OE series optical probes with a DPO70000 real-time scope deliver the performance and advanced debug capabilities designers need to fully troubleshoot 400G PAM4 signals, up to 56 GBaud, and reduce time to market.
These probes can also be used as a conventional O/E with flat response filter for general laser signal acquisition up to their respective bandwidth of 33 GHz or 59 GHz.
Model |
Wavelength Response |
High-bandwidth DC |
Optical Noise |
NRZ / PAM2 and PAM4 |
|
DPO7OE1 |
750nm - 1650nm
|
up to 33 GHz
|
6.6 uW rms (TekConnect / ATI)
|
up to 28 Gbaud
|
|
DPO7OE2 |
1200m - 1650nm
|
up to 59 GHz
|
10 uW rms (ATI)
|
up to 56 Gbaud
|
|
Power Rail Probes
Optimized for power integrity measurements, these probes combine low noise contribution, DC offset up to 60 V, high bandwidth, low loading, and a wide range of connectivity alternatives.
Model |
Attenuation |
Bandwidth |
Dynamic Range |
DC Offset |
|
TPR1000 |
1.25x
|
1 GHz
|
±1 V
|
±60 V
|
|
TPR4000 |
1.25x
|
4 GHz
|
±1 V
|
±60 V
|
|
Current Probe
Tektronix current probe solutions offer:
- The broadest range of AC/DC and AC-only current probes
- Measurement accuracy from uAs to 2000 A
- Best-in-class bandwidth up to 120 MHz
- Best-in-class current clamp sensitivity down to 1 mA
- The only products with 3rd Party Safety Certification (UL, CSA, ETL)
- The only products with bare wire voltage ratings
- Automatic readout and scaling when used with Tektronix oscilloscopes so you don’t have to convert volts to amps or manually set the scaling
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Model Maximum Current Minimum Current Bandwidth Rise Time
A621 RMS: 1000 A Peak: 2000 A 10 mA 50 kHz-
A622 DC: 100 A Max RMS: 71 A Peak Current: 100 A 50mA 100 kHz-
CT2 Peak: 36 A Max RMS: 2.5 A continuous mA 200 MHz < 500 ps
CT1 Peak: 12 A Max RMS: 450 mA continuous 200 μA 1 GHz < 350 ps
CT6 Peak: 6 A Max RMS: 120 mA continuous 200 μA 2 GHz < 200 ps
TRCP0300 300A 250 mA 30 kHz-
TCP305A DC: 50 A Max RMS: 35.4 APeak Current: 50 A 5 mA 50 kHz- < 7 ns
P6021A Max RMS: 10.6 A Peak Current: 250 A 10 mA 60 MHz- < 5.8 ns
TRCP3000 3000 A 500 mA 16 MHz
TRCP0600 600 A 500 mA 30 MHz
P6022 Max RMS: 4 A Peak Current: 100 A 1 mA 120 MHz- < 2.9 ns
TCP312A DC: 30 A Max RMS: 21.2 A Peak Current: 50 A 1 mA 100 MHz < 3.5 ns
TCP202A DC: 15 A Max RMS: 15 A Peak Current: 50 A 10mA 50MHz < 7 ns
TCPA400 (Probe Dependent) 100 MHz
|
|
|
|
|
|
TCP2020 20 A RMS, 100 A Peak Pulse 10 mA 50 MHz ≤ 7ns |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
TCPA300 DC - 100 MHz |
|
|
|
|
|
TCP0020 DC: 20 A Max RMS: 20 A Peak Current: 100 A 10 mA 50 MHz < 7 ns
TCP303 DC: 150 A Max RMS: 150 A Peak Current: 500 A 5 mA 15 MHz < 23 ns
TCP0030A DC: 30 A Max RMS: 30 A Peak Current: 50 A 1 mA 120 MHz < 2.92 ns
TCP404XL DC: 500 A Max RMS: 500 A Peak Current: 750 A 1 A 2 MHz < 175 ns
TCP0150 DC: 150 A Max RMS: 150 A Peak Current: 500 A 5 mA 20MHz < 17.5 ns
Differential Probe - High Voltage
A high voltage differential probe is used for measuring the voltage difference between two test points where neither test point is at ground. High voltage differential probes from Tektronix can be used for signals up to 6000 V. These probes are the best choice for making non-ground referenced, floating or isolated measurements in large part due to their common mode rejection capability. These products are designed, manufactured, and serviced by Tektronix.Tektronix High Voltage Differential Probe solutions offer:Best-in-class bandwidth and probe loadingThe only products with 3rd Party Safety Certification (UL, CSA, ETL)High and medium voltage products to support varying dynamic range and measurement resolution requirementsMost extensive set of probe accessories
|
|
|
|
|
|
MODEL Attenuation Bandwidth Input Impedance Maximum Voltage |
|
|
|
|
|
P5200A 50X / 500X 50 MHz 10 MΩ || 2 pF 500X:±1300 V 50X: ± 130 V
P5202A 20X / 200X 100 MHz 5 MΩ || 2 pF 200X: ± 640 V 20X: ± 64 V
P5205A 50X / 500X 100 MHz 10 MΩ || 2 pF 500X: ± 1300 V 50X: ± 130 V
THDP0200 50X / 500X 200 MHz 10 MΩ || 2 pF 500X: ± 1500 V 50X: ± 150 V
TMDP0200 25X / 250X 200 MHz 5 MΩ || 2 pF 250X: ± 750 V 25X: ± 75 V
P5210A 100X / 1000X 50 MHz 40 MΩ || 2.5 pF 1000X: ± 5600 V 100X: ± 560 V
THDP0100 100X / 1000X 100 MHz 40 MΩ || 2.5 pF 1000X: ± 6000 V 100X: ± 600 V
TDP0500 5X / 50X 500 MHz 1 MΩ || ≤ 1 pF 50X: ± 42 V 5X: ± 4.2 V
TDP1000 5X / 50X 1 GHz 1 MΩ || ≤ 1 pF 50X: ± 42 V 5X: ± 4.2 V
P6251 5X / 50X 1 GHz 1 MΩ || ≤ 1 pF 50X: ± 42 V 5X: ± 4.2 V
Differential Probe - Low Voltage
Differential signaling used in high speed serial standards requires very accurate characterization. The industry-leading bandwidth and signal fidelity found in a Tektronix low voltage differential probe ensures that you see every possible detail. Tektronix offers TriMode™ architecture which streamlines measurement acquisition by enabling you to make differential, single-ended, and common mode measurements with a single connection!
MODEL Attenuation Bandwidth Input Impedance Maximum Voltage
ADA400A 1 MHz 1 MΩ || ~ 55 pF 100x, 10x, 1x, 0.1x
TDP1500 1.5 GHz 200 kΩ / ≤1 pF 1X / 10X ±25 V (DC + pk AC)
P6247 1 GHz 200 kΩ / < 1 pF 1X, 10X ±25 V (DC + pk AC)
TDP3500 3.5 GHz 100 kΩ / ≤0.3 pF 5X ±15 V (DC + pk AC)
TDP4000 4 GHz 100 kΩ || ≤ 0.3 pF 5X ±15 V (DC + pk AC)
P7504 4 GHz - 5X / 12.5X ±15 V (DC + pk AC)
P6248 1.5 GHz 200 kΩ / <1 pF 1X, 10X ±25 V (DC + pk AC)
TDP7704 4 GHz 100kΩ (SIT) / 150kΩ (BRW) || 0.4 pF 4X / 10X ±12 V
TDP7706 6 GHz 100kΩ (SIT) / 150kΩ (BRW) || 0.4 pF 4X / 10X ±12 V
TDP7708 8 GHz 100kΩ (SIT) / 150kΩ (BRW) || 0.4 pF 4X / 10X ±12 V
P7506 6 GHz 100 kΩ 5X / 12.5X ±15 V (DC + pk AC)
P7708 8 GHz 100kΩ (SIT) / 150kΩ (BRW) / 0.4 pF 4X / 10X ±12 V
P7508 8 GHz 100 kΩ 5X / 12.5X ±15 V (DC + pk AC)
P7713 13 GHz 100kΩ (SIT) / 150kΩ (BRW) / 0.4 pF 4X / 10X ±12 V
P7513A >13 GHz 100 kΩ5X / 12.5X ±15 V (DC + pk AC)
P7716 16 GHz 100kΩ (SIT) / 150kΩ (BRW) / 0.4 pF 4X / 10X ±12 V
P7516 16 GHz - 5X / 12.5X ±15 V (DC + pk AC)
P7625 25 GHz 50 Ω / 225 Ω0 25/0.5/1/2/4 (coax)1.25/2.5/5/10/20 (P76TA) ±5 V (DC + pk AC)
P7720 20 GHz 100kΩ (SIT) / 150kΩ (BRW) / 0.4 pF 4X / 10X ±12 V
P7520A 20 GHz - 5X / 12.5X ±15 V (DC + pk AC)U
P7633 33 GHz 50 Ω / 225 Ω 0.25/0.5/1/2/4 (coax)1.25/2.5/5/10/20 (P76TA) ±5 V (DC + pk AC)
Passive Probe
Passive voltage probes ship standard with most oscilloscopes and provide a low cost, general purpose probing solution. Generally, these probes lack the performance of an active voltage probe but provide the ruggedness and wide dynamic range suitable for visualizing signals over a broad range of applications. Tektronix has released a new class of passive probes that redefine performance in the passive probe product category.Tektronix new class of passive probe solutions offer:Best-in-class bandwidth up to 1 GHzBest-in-class input capacitance as low as 3.9 pF which minimizes probe loading effectsBest-in-class input capacitance which minimizes performance loss when long ground leads are attachedAutomated probe compensation eliminating the need for the compensation screwdriver
MODEL Attenuation Bandwidth Input Impedance Maximum Voltage |
|
|
TPP0051 10x 50 MHz 10 MΩ || 12 pF 300 Vrms CAT II
TPP0100 10X 100 MHz 10 MΩ || 12 pF 300 Vrms CAT II
TPP0101 10X 100 MHz 10 MΩ || 12 pF 300 Vrms CAT II
TPP0200 10X 200 MHz 10 MΩ || 12 pF 300 Vrms CAT II
TPP0201 10X 200 MHz 10 MΩ || 12 pF 300 Vrms CAT II
P2220 1X / 10X 10X:200 MHz 1X: 6 MHz 10X: 10MOhm|| 17pF,1X: 1MOhm|| 110pF 10X: 300 Vrms CAT II,1X: 150 Vrms CAT II
P2221 1X / 10X 10X: 200 MHz1X: 6 MHz 10X: 10 MΩ || 110 pF 1X: 1 MΩ || 17 pF 10X: 300 Vrms CAT II1X: 150 Vrms CAT II
P6101B 1X 15 MHz 1 MΩ || 100 pF 300 Vrms CAT II
P3010 10X 100 MHz 10 MΩ || 13 pF 300 Vrms CAT II
TPP0250 10X 250 MHz 10 MΩ || 3.9 pF 300 Vrms CAT II
TPP0500B 10X 500 MHz 10 MΩ || 3.9 pF 300 Vrms CAT
TPP0502 2X 500 MHz 2 MΩ || 12.7 pF 300 Vrms CAT II
P6139B 10X 500 MHz 10 MΩ || 8 pF 300 Vrms CAT II
P5050B 10X 500 MHz 10 MΩ || 11 pF 300 Vrms CAT II
TPP1000 10X 1 GHz 10 MΩ || 3.9 pF 300 Vrms CAT II
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
High Voltage Probe - Single Ended
A high voltage single-ended probe is typically used for measuring ground referenced signals up to 40 kV. However, some single-ended probes are designed for instruments with isolated or floating inputs for measurements that are not ground referenced. Users should select probes with a low input capacitance specification (< 4 pF) to minimize the probe’s loading effect on the circuit because a probe with lower input capacitance will offer higher input impedance at higher frequencies.Tektronix High Voltage Probe solutions offer:Best-in-class bandwidth up to 800 MHzBest-in-class probe loading with input capacitance as low as 1.8 pFThe only products with 3rd Party Safety Certification (UL, CSA, ETL)Most extensive set of probe accessories
MODEL Bandwidth Input Impedance Maximum Voltage |
|
|
|
|
|
P5122 200 MHz 100 MΩ || 4 pF RMS: 1000 V CAT II
P5100A 500 MHz 40 MΩ || 2.5 p Peak: 2.5 kV RMS: 1000 V CAT II
P5150 500 MHz 40 MΩ || 3.8 pF Peak: 2.5 kVRMS: 1000 V CAT II
TPP0850 800 MHz 40 MΩ || 1.8 pF Peak: 2.5 kV RMS: 1000 V CAT II
P6015A 75 MHz 100 MΩ || 3 pF 20 kV
Low Voltage Probe - Single Ended
A low voltage single-ended probe is typically used for measuring high-speed, ground referenced signals up to 12 V. These low voltage probes are the best choice for making measurements on high impedance, high frequency circuit elements which require minimal probe loading. Users should select probes with a low input capacitance specification (~1 pF) to minimize the probe’s loading effect on the circuit. A probe with lower input capacitance will offer higher input impedance at higher frequencies.Tektronix Low Voltage Single-ended Probe solutions offer:Bandwidths up to 4 GHz.Very high input impedance with low input capacitance (<1 pF)Most extensive set of probe accessories for optimum measurement performance
|
|
|
|
|
|
MODEL Attenuation Bandwidth Input Impedance Maximum Voltage |
|
|
|
|
|
|
|
|
|
|
|
P6243 10X 1 GHz 1 MΩ || ≤ 1 pF ± 8 V
P6245 10X 1.5 GHz 1 MΩ || ≤ 1 pF ± 8 V |
|
|
|
|
|
|
|
|
|
|
|
TAP1500 10X 1.5 GHz 1MΩ || ≤ 1 pF ± 8 V |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
TAP2500 10X 2.5 GHz 40 kΩ || ≤ 0.8 pF ± 4 V
TAP3500 10X 3.5 GHz 40 kΩ || ≤ 0.8 pF ± 4 V
TAP4000 10X 4 GHz 40 kΩ || ≤ 0.8 pF ± 4 V
|
|
|
|
|
|
|
|
|
|
|
|
P7240 5X 4 GHz 20 kΩ || ≤ 0.8 pF ± 2 V
Logic Analyzer Probe
Verification and debug of today's high speed, low voltage digital signals requires probing solutions that can accurately acquire from a wide variety of electronic designs and protect signal fidelity. Tektronix logic analyzer probes contain a variety of connectivity options that are engineered to ensure that signal acquisition is a true reflection of your design's performance. Compare and learn more about Tektronix logic analyzer probe solutions.
Model |
Connectivity |
Instrument Series |
|
P5910 |
Fits both 0.100 in. and 2 mm square pin configurations |
TLA6400 |
|
TLP058 |
FlexChannel (TM) input |
5 Series MSO |
|
P5934 |
Amp Mictor 34-channel connector |
TLA6400 |
|
P5960 |
D-Max® probing technology compression cLGA |
TLA6400 |
|
Low Bandwidth Optical Probes
Depend on Tektronix to provide you with performance you can count on, and industry-leading service and support.Broad Wavelength Response 500 to 950 nm or 1100 to 1700 nmHigh-bandwidth DC up to 1.2 GHzHigh Gain 1 V/mWLow Noise <11 pW/√HzProbe Connects to DPO7000*1 and DPO/DSA/MSO70000*2SeriesSONET/SDH and Fibre Channel Reference Receiver Performance
Model |
Bandwidth |
|
P6703B |
≥1.2 GHz |
|
P6701B |
≥1.0 GHz |
|